The MR810 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the MR810 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the MR810 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output | Separation distance according to the frequency of | ||
|
| transmitter (m) |
|
power of transmitter | 150 kHz to 80 MHz | 80 to 800 MHz | 800 MHz to 2,5 GHz |
(W) | d = 1,2√P | d = 1,2√P | d = 2,3√P |
| |||
0,01 | 0,12 | 0,12 | 0,23 |
0,1 | 0,38 | 0,38 | 0,73 |
1 | 1,2 | 1,2 | 2,3 |
10 | 3,8 | 3,8 | 7,3 |
100 | 12 | 12 | 23 |
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance of the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
D
MR810 Technical Manual Revision C REF 185042601 | 43 |