Appendix D: ELECTROMAGNETIC COMPATIBILITY

D.1 COMPLIANCE AND WARNINGS

The MR810 with RT308 breathing circuit has been tested to and found compliant with the requirements of IEC 60601-1-2: 2001.

WARNING:

The use of other accessories other than those specified may result in increased emissions or decreased immunity of the equipment.

Verify correct operation if the MR810 is to be used adjacent to other equipment.

D.2 ELECTROMAGNETIC EMISSIONS
Guidance and manufacturer’s declaration – electromagnetic emissions

The MR810 is intended for use in the electromagnetic environment specified below. The customer or the user of the MR810 should assure that it is used in such an environment.

Emissions TestComplianceElectromagnetic environment - guidance

RF emissions

Group 1

The MR810 uses RF energy only for its internal

CISPR 11

 

function. Therefore, its RF emissions are very low

 

 

and not likely to cause any interference in nearby

 

 

electronic equipment.

RF emissions

Class A

The MR810 is suitable for use in all establishments

Harmonic emissions

Class A

other than domestic and those directly connected to

IEC 61000-3-2

 

the public low-voltage power supply network that

Voltage fluctuations/

Complies

supplies buildings used for domestic purposes.

flicker emissions

 

 

IEC 61000-3-3

 

 

D

40

MR810 Technical Manual Rev C REF 185042601