71
Remote Operation
Interface Commands
Command Function
Associated Setting or
Function (screen)
CAL:DATE Sets the instrument’s date of calibration Calibration Date (Calibration
Parameters)
CAL:PAR:LIN<n> Sets a calibration linearity correction parameter Linearity C<n> (Calibration
Parameters)
CAL:PAR:RES Sets the calibrated resistance of an internal
reference resistor
W Resistor (Calibration
Parameters)
CAL:PAR:ZERO Sets the calibration zero correction parameter Zero Correction (Calibration
Parameters)
CALC<n>:AVER:ALL? Reads a channel’s statistical values (Main Screen)
CALC<n>:AVER:AVER? Reads a channel’s statistical mean Mean (Main Screen)
CALC<n>:AVER:CLEA Clears a channel’s statistics RESET STATISTICS (Main Screen)
CALC:AVER:COUN Sets the statistics Sample Count setting Sample Count (Statistics Settings)
CALC<n>:AVER:MAX? Reads a channel’s statistical maximum Maximum (Main Screen)
CALC<n>:AVER:MIN? Reads a channel’s statistical minimum Minimum (Main Screen)
CALC<n>:AVER:SDEV? Reads a channel’s statistical sample standard
deviation
Std Dev (Main Screen)
CALC<n>:AVER:SEM? Reads a channel’s statistical standard error of the
mean
SE of Mean (Main Screen)
CALC<n>:AVER:SPR? Reads a channel’s statistical spread Spread (Main Screen)
CALC<n>:AVER:SSIZ? Reads a channel’s statistical sample size N (Main Screen)
CALC:AVER:TCON Sets the Statistics Mode setting Statistics Mode (Statistics Settings)
CALC:DELT:CHAN Sets the Delta Channel setting Delta Channel (Statistics Settings)
CALC<n>:DELT:DATA? Reads a channel’s delta result Delta (Main Screen)
CALC:DELT:MODE Sets the Delta Mode setting Delta Mode (Statistics Settings)
CALC:DELT:OFFS Sets the Delta Offset setting Delta Offset (Statistics Settings)
CALC<n>:TYPE Sets the channel’s Calculation setting Calculation (Channel Settings)
DATA:REC:ALL Sets the data recording Record Statistics setting Record Statistics (Recording
Settings)
DATA:REC:FILE Sets the data recording USB File setting USB File (Recording Settings)
DATA:REC:MEM Sets the data recording Memory Device setting Memory Device (Recording Settings)
DATA:REC[:STAT] Enables data recording START/STOP RECORDING (Main
Screen)
DISP:ALER:CLEA? Clears any alert message that is active CLEAR MESSAGE (Main Screen)
DISP:ALER[:READ]? Reads the display alert message (Main Screen)
DISP:DATE:FORM Sets the display Date Format setting Date Format (Display Settings)
DISP:DEC:FORM Sets the display Decimal setting Decimal (Display Settings)
DISP:FIEL:FEED<n>? Reads a statistics eld setting of the visible statistics
elds memory set
(Main Screen)
DISP:FIEL:MEM[?] <n> Reads or sets the statistics elds memory set for
display
Recall Memory (Field Settings)
DISP:FIEL:MEM<n>:FEED
<n>[?] <s>
Reads or sets a statistics eld setting of a statistics
elds memory set
Field <n> (Field Settings)
DISP:LANG Sets the display Language setting Language (Display Settings)
DISP:TEMP:DIG Sets the display Temperature Resolution setting Temperature Resolution (Display
Settings)
DISP:WARN:DUE Sets the display Rs Due Warning enable Rs Due Warning (Display Settings)
DISP:WARN:ITS Sets the display ITS-90 Subrange Warning enable ITS-90 Subrange Warning (Display
Settings)
FETC? Reads a channel’s latest measurement