1594A/1595A Super-Thermometer
Interface Commands
82
Reads a statistics eld setting of the visible statistics elds memory set (see Section 5.7.4.2, FIELD SET-
TINGS (F2), on page 48).
Sufx <n> is the eld number, which may be 1 through 4. The response will be one of the following: MEAN,
SDEV, SEM, MAX, MIN, SPR, N, DELT, RES, RAT, TIME, or DATE.
8.7.3.37 DISP:FIEL:MEM <n>
Example command: DISP:FIEL:MEM 4
Query form: DISP:FIEL:MEM<n>?
Alternate parameter: MIN, MAX, DEF
Reads or sets a statistics elds memory set for display (see Section 5.7.4.2, FIELD SETTINGS (F2), on page
48).
The parameter <n> may be a number from 1 to 4.
8.7.3.38 DISP:FIEL:MEM<n1>:FEED<n2> <feed>
Example command: DISP:FIEL:MEM4:FEED4 TIME
Query form: DISP:FIEL:MEM<n1>:FEED<n2>?
Alternate parameter: DEF
Reads or sets a statistics eld setting of a statistics elds memory set (see Section 5.7.4.2, FIELD SETTINGS
(F2), on page 48).
Sufx <n1> is the memory set number, which may be 1 through 4. Sufx <n2> is the eld number, which may
be 1 through 4. The parameter <feed> may be one of the following: MEAN, SDEV, SEM, MAX, MIN, SPR,
N, DELT, RES, RAT, TIME, or DATE.
8.7.3.39 DISP:HELP <boolean>
Query form: DISP:HELP?
Sets the display Start-up Help Screen enable.
This setting determines whether the Start-up Help Screen appears at power-on. By default it is 1 and the Start-
up Help Screen appears each time the instrument is powered on. If the setting is changed to 0, the Start-up
Help Screen will no longer appear.
8.7.3.40 DISP:LANG ENGLISH|CHINESE|JAPANESE|SPANISH|GERMAN|FRENCH|RUSSIAN
Query form: DISP:LANG?
Alternate parameter: DEF
Sets the display Language setting (see Section 5.7.4.1, USER SETTINGS (F1), on page 47). The default
setting is ENGLISH.
8.7.3.41 DISP:TEMP:DIG <digits>
Query form: DISP:TEMP:DIG?
Alternate parameter: DEF, MIN, MAX
Sets the display Temperature Resolution setting (see Section 5.7.4.1, USER SETTINGS (F1), on page 47).
Parameter <digits> species the number of decimal places. The range is 1 to 6 and the default is 5.
8.7.3.42 DISP:WARN:DUE <boolean>
Query form: DISP:WARN:DUE?
Alternate parameter: DEF