4g) Remove the Light Wand and replace it on the same point on the biceps. Press GREEN BUTTON on the Light Wand. The display will flash a decimal point, then the number “1b”.

4h) Place Light Wand firmly on Triceps of same arm (see page 16 for exact measurement point).

NOTE: Aim silver stripe directly toward the shoulder. Keep arm relaxed.

Press ENTER. The display will flash a decimal point. Don’t move the Light Wand until “2b” is displayed.

4i) Remove Light Wand and replace it on the same point on the triceps. Press ENTER. The display will flash a decimal point. Don’t move Light Wand until “1c” is displayed.

4j) Place Light Wand firmly on thigh at same side of body that the arm was measured; e.g. if right arm, then measure right thigh. (See page 16 for exact measurement point.)

NOTE: Aim silver stripe directly up the leg. Press ENTER. The display will flash a decimal point. Don’t move Light Wand until “2c” is displayed.

4k) Remove Light Wand and replace it on the same point on the thigh. Press ENTER. Display will again flash a decimal point. Don’t move the Light Wand until percent fat is displayed and printing starts.

NOTE: Steps 5 through 16 are not applicable. Those steps are for measurements using only the biceps.