PROBLEM

PROBABLE CAUSE

CORRECTIVE MEASURES

3) Bad Repeatability

a) Inconsistent Measurement Technique

a) Ensure that there us enough pressure

 

 

applied between the Light Wand and the

 

b) Incorrect Zero Adjustment Procedure

subject’s arm

 

 

 

c) Not measuring at the proper location

b) When taking a measurement, ensure

 

that the Silver Stripe on the Light Shield is

 

on the arm

pointing towards the subject’s shoulder

 

 

(the Silver Stripe on the Light Wand

 

 

should be aligned with the Silver Stripe on

 

 

the Light Shield).

 

 

c) Always use the Light Shield to block

 

 

external light.

 

 

d) Avoid taking measurements under

 

 

strong lights (e.g. direct sunlight) as errors

 

 

may occur

 

 

e) When Zero Adjusting, ensure that the

 

 

silver line on the Light Wand is aligned

 

 

with the silver line on the Optical

 

 

Standard.

 

 

f) When Zero Adjusting, make sure that

 

 

the Light Shield has been removed from

 

 

the Light Wand. This is to ensure that the

 

 

Light Wand is properly seated in the

 

 

Optical Standard.

 

 

g) Always use the Biceps Locator to find

 

 

the measurement site. Instructions can

 

 

be found on page 12.