TROUBLESHOOTING

TROUBLESHOOTING

PROBLEM

PROBABLE CAUSE

CORRECTIVE MEASURES

1) Display shows "LOBAT"

Batteries are low

a) Replace batteries (Refer to page 3).

 

a) Bad connection between Light Wand

a) Be sure the Light Wand is properly

2) Display shows "1999.9" or

and Probe Socket

inserted into the Probe Socket.

"-1999.99"

 

 

 

b) Too much/too little light

b) Be sure the Light Wand is properly

 

 

seated in the Optical Standard when

 

 

Zero Adjusting.

3) Bad Repeatability

a) Inconsistent measurement technique

a) Make sure you are applying enough

 

 

pressure between the Light Wand and

 

 

the subject's arm.

 

 

b) The silver line on the Light Wand

 

 

must be aligned with the line on the

 

 

Optical Standard when Zero Adjusting.

 

 

c) When taking a measurement, the

 

 

silver line on the Light Wand must point

 

 

toward the subject's shoulder.

 

 

d) Use the Light Shield to block the

 

 

external light. Always avoid

 

 

measurements under strong lights (e.g.

 

 

direct sunlight) because errors may

 

 

occur.