PROBLEM

PROBABLE CAUSE

CORRECTIVE MEASURES

4) Unusually high body fat estimates

a) Too much light present during the

a) Do not lift the Light Wand from the

(40-50%)

measurement.

subject's biceps until a "2" appears on

 

 

the screen during the first estimation or

 

 

the printer starts printing after the

 

 

second estimation.

 

 

b) Use the Light Shield to block the

 

 

external light. Always avoid

 

 

measurements under strong lights (e.g.

 

 

direct sunlight) because errors may

 

 

occur.

5) Printer Stops

a) Low batteries.

a) Replace batteries (see page 3).

 

b) Bad battery contact.

b) Remove batteries and clean tabs with

 

 

isopropyl alcohol (rubbing alcohol).

 

c) Bad AC connection.

 

 

 

c) Unplug the AC Adapter from the wall

 

 

socket and from the unit. Clean the

 

 

adapter contact with isopropyl alcohol

 

 

(rubbing alcohol).

30

Page 31
Image 31
Futrex Futrex -5000/XL manual Problem Probable Cause Corrective Measures