Chapter 3: Getting Started

The maximum circuit load for the electrical relay is 6 amps. Only the remote blower wiring is connected inside the electrical junction box. The lights and airflow monitor are powered from a standard duplex electrical receptacle located nearby for connecting the power cord. All wiring for the XPert Filtered Balance Station should be performed by a licensed electrician and conform to local codes. In most cases, the XPert Filtered Balance Station will require the use of shielded conduit to protect the wiring. See Figure 3-4. The grounding connection shall not be made to the terminal box cover. The building electrical supply system for the remote blower should include overload protection such as a switch or circuit breaker in close proximity and within easy reach of the operator. The switch or circuit breaker shall be marked as the disconnecting device. Consult NEC-2002 for proper installation.