SEFA 1-2002

Fume hood face velocities based on toxicity levels of chemicals

Class A – 125 to 150 fpm Class B – 80 to100 fpm Class C – 75-to 80 fpm

Test method – face velocity profile and smoke generation

Scientific Equipment & Furniture Association

1028 Duchess Drive

McLean, VA 22102

(703)538-6007 www.sefalabs.com

NFPA 45 – 2002 Fire Protection for Laboratories Using

Chemicals

Laboratory hoods should not be relied on for explosion protection

Exhaust air from fume hoods should not be recirculated

Services should be external to the hood

Canopy hoods only for non-hazardous applications

Materials of construction should have flame spread of 25 or less

80 to 120 fpm to prevent escape

NFPA 30 – 2000 Flammable and Combustible Liquids

Code

Approved cabinets may be metal or wood

Vent location on cabinets are required

Venting of cabinets not a requirement

National Fire Protection Association

1 Batterymarch Park

P.O. Box 9101

Quincy, MA 02269-9101

(800)344-3555 www.nfpa.org