Chapter 6: Maintaining Your Filtered Enclosure

To start the calibration, use the Labconco 1000 Manual and enter the “CALIBRATIION” mode on the display. Follow the instructions on the display and alter the low and high exhaust volumes with the speed control. The following suggested inflow face velocity speeds are recommended to successfully calibrate. Typical low air alarms are set 10-30 fpm below operational speeds. Typical yellow cautions are set 3-10 fpm above the red low air alarm. Be sure to enter “Done” after completing the calibration in “SETUP.” Once completed, go to “RUN” and hit “ENTER.”

 

Nominal Enclosure

Low

High

Low Air Alarm

Operating In Flow

Calibration

Calibration

Set Point

Speed

Set Point

Set Point

50 fpm

60 fpm

50-60 fpm

100-120 fpm

60-70 fpm

80 fpm

50-70 fpm

100-140 fpm

70-90 fpm

100 fpm

60-80 fpm

110-140 fpm

 

 

 

 

Factory Settings

 

 

 

60 ± 10 fpm

90 ± 10 fpm

60 ± 10 fpm

90 ± 10 fpm

54

Page 59
Image 59
Gentek 3940421, 3940200, 3950320, 3950202, 3940320, 3940402, 3950222, 3940300, 3940322, 3950221, 3950220, 3940222 Factory Settings