CHAPTER 8

TROUBLESHOOTING AND

SERVICER OPERATING

LOG

Refer to the following table if your filtered enclosure fails to operate properly. If the suggested corrective actions do not solve your problem, contact Labconco for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

 

 

 

Contaminants outside

Improper user

See “Certifying the Enclosure” Chapter

of enclosure.

techniques for the

3 and “Safety Precautions” Chapter 4

 

enclosure.

sections in the manual. (Ref. Appendix

 

 

D).

 

Restriction of the

Remove obstruction to ensure that all

 

baffle air slots or

air slots and the exhaust outlet are

 

blockage of the

unobstructed.

 

exhaust outlet.

 

 

External factors are

See “Location Requirements” Chapter

 

disrupting the

2, “Certifying the Enclosure” Chapter

 

enclosure airflow

3, and “Safety Precautions” Chapter 4

 

patterns or acting as a

sections of this manual. (Ref. Appendix

 

source of

D).

 

contamination.

 

 

Enclosure has

Have enclosure certified and check

 

improper face

exhaust system. Check HEPA filters for

 

velocity.

loading. Adjust enclosure speed

 

 

control. Enclosure should have an

 

 

average face velocity of 60-100 fpm for

 

 

weighing procedures and 75-105 fpm

 

 

for biohazardous procedures.

69

Page 74
Image 74