Chapter 8: Troubleshooting

PROBLEM

CAUSE

CORRECTIVE ACTION

Airflow Monitor

Wrong alarm set

Airflow monitor was not properly

Malfunction

point.

adjusted. Repeat calibration steps

(Continued)

 

outlined in this manual in Chapter 6

 

 

or in the supplied Digital 1000

 

 

Monitor Manual.

 

Constant audible

Check airflow and calibration of

 

alarm.

airflow monitor. See Chapter 6.

 

Continuous alarm.

Check the face velocity of the

 

 

enclosure as the airflow of the

 

 

system may have changed. If

 

 

incorrect, adjust the speed control to

 

 

increase face velocity. The HEPA

 

 

filter may have become loaded. If

 

 

face velocity is correct, calibrate the

 

 

airflow monitor outlined in this

 

 

manual in Chapter 6.

 

Monitor alarms; air

Lightly clean the airway with clean

 

way to airflow

air. Be careful not to touch sensitive

 

monitor sensor is

electrical components.

 

blocked by insects,

 

 

dust or debris.

 

 

Audible disable

An alarm condition must be

 

will not stay

continuously present before the

 

operational.

audible alarm can be silenced. If

 

 

flow conditions fluctuate near the

 

 

alarm set point, the airflow monitor

 

 

will automatically reset it. Action

 

 

should be taken to bring the

 

 

enclosure airflow into proper

 

 

operating parameters or adjust the

 

 

alarm set point lower.

71