Scale Calibration Procedure (lbs. or kg) with Remote Display
1. Temporarily Remove the platter and place it in a safe location.
Note: | It is the responsibility of the owner of the scale to confirm compliance with the relevant Weights |
| and Measures statutes and regulations applicable in your area by checking with the appropriate |
| government agency before placing a newly calibrated unit into service or removing any official |
| seals. |
2.If this is a currently installed scanner/scale in need of calibration, cut and remove the calibration switch cover seal. If this is a new installation, cut and remove the
Follow all
3.Remove the M3 screw securing the calibration switch/button cover. Place the cover and screw in a safe location. Verify that the scale calibration switch is in the Run position.
Figure 38. Calibration Switch/Button Cover Removal