Scale Calibration Procedure (lbs. or kg) with Remote Display
7.Make sure there is no load on the scale platter.
8.Calibrate a zero load. Press the right arrow (►) button once when the remote display reads CAL 1 and there is no load on the scale platter.
Figure 42.
9.Place a half load (15.00 lbs. or 7.500 kg) on the center of the scale platter then press the right arrow (►) button.
Figure 43.
10.Add an additional half load (15.00 lbs. or 7.500 kg) to the existing half to simulate a full load, center entire load then press the right arrow (►) button.
Figure 44.
11.Remove half of the load (15.00 lbs. or 7.500 kg), center the remaining load then press the right arrow ( ►) button.
Figure 45.
48