6.8 Setting Range Values Using Local Adjustments,
Continued

Procedure, continued

Table 27 Setting Range Values Using Local Zero and Span

Adjustments, continued

Step Action

10 Wait 30 seconds so that changes have been copied to the
transmitter’s non-volatile memory.

11

Remove applied pressure and turn OFF transmitter power.
12 Replace end-cap on PWA side of electronics housing and tighten
lock.
13 Remove milliammeter from TEST terminals and replace end-cap and
tighten lock.
14 Turn ON transmitter power and check Local Smart Meter reading, if
applicable.

Figure 32 Typical Setup for Setting Range Values Using Local Zero and Span Adjustments.

+
-
Field
Terminals
- SIGNAL +
+ -
TEST
Power
Supply
+
-
Receiver

ST 3000

250
Ω
STR3015
Milliammeter