6/08 ST 3000 Release 300 and SFC Model STS103 User’s Manual 105
6.12 Configuring Smart Meter Using Pushbuttons, Continued
Transmitter Output
Conformity and Smart
Meter Configuration,
continued

Table 32 Smart Meter Engineering Units Code

Smart Meter Code Engineering Unit Transmitter Output
Conformity
EU0
% * Linear or Square Root
EU1
in H2O *
EU2 mmHg *
EU3 PSI *
EU4 kPa
EU5 MPa
EU6 mbar Linear
EU7 bar
EU8 g/cm2
EU9 kg/cm2
EUA mmH2O
EUB inHg
EUC mH2O
EUD GPM * Square Root
EUE GPH * Square Root
EUF Custom Linear or Square Root
* These selections have indicators on smart meter display.
Use stick-on labels provided for other engineering units.
Selecting Engineering
Units The procedure in Table 33 outlines the steps for selecting the desired

engineering units for a Local Smart Meter using its local adjustments on

the face of the meter. You will be selecting the unit of measurement

that you want the smart meter to indicate during normal operation.

WARNING When the transmitter’s end-cap is removed, the housing is not

explosionproof.

Continued on next page