ST 3000 Release 300 and SFC Model STS103 User’s Manual 6/08
68
6.2 Overview, Continued
Configuration
decision summary,
continued

Figure 25 Flowchart — ST 3000 Pressure Transmitter Configuration, continued.

Do you know what
range values to use
?

NO

YES

Deter min e Lower Ra ng e Val ue (LRV) and Upper Ra nge
Value (URV) to be used for setting range values.
Set LRV and URV to desir ed
values through SFC prom p ts
LRV 1 and URV 1.
See procedure in Table 25 or 26 fo r range setting deta ils.
Use [LRV] and [URV] keys to ca l l up resp ecti ve SFC prom pts
for keying in range values or setting range v alues to applied
pressures.
Will transmitter be
operat ing in DE
mode?

NO

YES

Sel ec t app ropr iat e DE
configuration parameters
thr ough SFC prompt DE CONF.
See procedures in Tables 28 and 29 for DE
para m ete r s ele ction s .
Pr e ss [ SHIFT] and [MENU ITEM] keys to call up
DE configuration menu.

C

B

continued on next page