Description of Diaphragm Seals
Diaphragm seals are traditionally used when a standard pressure transmitter should not be exposed to the process pressure directly. Diaphragm seals typically protect the pressure transmitter from one or more damaging aspects of the process media. Consideration for using a diaphragm seal should be made in the following circumstances.
•High Process Temperature
•Process Media is Viscous or Contains Suspended Solids
•Process Media is Subject to Solidifying
•Process Media is Corrosive
•Process Application Requires Sanitary Connections
•Process Application Subjects the Measuring Instrument to Hydrogen Permeation
•Tank Level Applications with Maintenance Intensive Wet Legs
•Tank Application with Density or Interface Measurements
•Measuring Instrument Requires Remote Mounting
The following diaphragm seals are standard from Honeywell (please call your local salesperson if you do not see the product you need for your application):
Figure 2 - Flush Flange Seals can be used with differential, gauge and absolute pressure transmitters and are available with 3” ANSI Class 150, ANSI Class 300 and DIN
Figure 2
Figure 3 - Flange Seal with Extended Diaphragm can be used with differential, gauge and absolute pressure transmitters and are available with 3” and 4” ANSI Class 150, ANSI Class 300, DIN
Figure 3
Figure 4 - Pancake Seals can be used with differential, gauge and absolute pressure transmitters and are available with 3” ANSI Class 150, 300 and 600 process connections.
Figure 4
Figure 5 - Chemical Tee “Taylor” Wedge seals can be used with differential pressure transmitters and are available with Taylor Wedge 5” O.D. process connection.
Figure 5