Honeywell STR12D warranty Specifications, Operating Conditions - All Models, 34-ST-03-64, Page

Models: STR12D

1 36
Download 36 pages 39.95 Kb
Page 5
Image 5
34-ST-03-64

 

 

 

 

 

 

 

 

 

 

 

 

 

34-ST-03-64

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Page 5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Specifications

 

 

 

 

 

 

 

Operating Conditions – All Models

 

 

 

 

 

 

 

 

 

 

 

Parameter

 

Reference Condition

Rated Condition

 

Operative Limits

Transportation and

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Storage

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

°C

 

°F

°C

 

°F

 

°C

 

°F

°C

°F

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Ambient Temperature *

25 ±1

 

77 ±2

 

 

 

–55 to 90

–67 to 194

 

 

Humidity

% RH

10 to 55

 

0 to 100

 

0 to 100

 

0 to 100

 

 

 

 

 

 

 

 

 

 

Maximum Allowable

 

MAWP is minimum of Body Rating or Seal Rating (See Model Selection Guide for Seal

 

 

Working Pressure (MAWP)

MAWP)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Body

 

MAWP

 

 

 

 

 

 

 

 

 

 

STR12D

 

2500 psig (172 bar) Bolted Process Heads Table I _ _ A

 

 

 

 

 

STR13D

 

2500 psig (172 bar) Bolted Process Heads Table I _ _ A

 

 

 

 

 

STR12D

 

1450 psig (100 bar) All Welded Process Heads Table I _ _ C

 

 

 

 

 

STR13D

 

1450 psig (100 bar) All Welded Process Heads Table I _ _ C

 

 

 

 

 

STR14G

 

500 psig (35 bar)

 

 

 

 

 

 

 

 

 

 

STR17G

 

3000psig (207 bar)

 

 

 

 

 

 

 

 

STR14A

 

500 psia (35 bar).

 

 

 

 

 

 

 

Vacuum Region - Minimum

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

mmHg absolute

 

Atmospheric (See Figure 15 for vacuum limitations.)

 

 

 

 

 

 

 

 

 

 

 

 

 

Supply Voltage, Current,

Voltage Range:

10.8 to 42.4 Vdc at terminals

 

 

 

 

 

and Load Resistance

 

Current Range:

3.0 to 21.8 mA

 

 

 

 

 

 

 

 

 

Load Resistance:

0 to 1440 ohms (as shown in Figure 16)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

* Ambient Temperature Limit is a function of Process Interface Temperature. (See Figure 13.)