| Page 15 | |
Table 2 AC Voltage Test: |
|
|
|
|
| ||
|
|
|
|
|
|
|
|
| Step | Range | Input | Frequency | Reading | Reading |
|
|
|
|
|
|
|
|
|
| 1 | 400mV | 300mV | 60 | 295.6 to304.4 | 295.6 to304.4 |
|
|
|
|
|
|
|
|
|
| 2 | 400mV | 300mV | 400 | 295.6 to304.4 | 295.6 to304.4 |
|
|
|
|
|
|
|
|
|
| 3 | 4V | 3.000V | 60Hz | 2.965 to 3.035 | 2.965 to 3.035 |
|
|
|
|
|
|
|
|
|
| 4 | 4V | 3.000V | 400Hz | 2.965 to 3.035 | 2.965 to 3.035 |
|
|
|
|
|
|
|
|
|
| 5 | 40V | 30.00V | 60Hz | 29.65 to 30.35 | 29.65 to 30.35 |
|
|
|
|
|
|
|
|
|
| 6 | 40V | 30.00V | 400Hz | 29.65 to 30.35 | 29.65 to 30.35 |
|
|
|
|
|
|
|
|
|
| 7 | 400V | 300.0V | 60Hz | 296.5 to 303.5 | 296.5 to 303.5 |
|
|
|
|
|
|
|
|
|
| 8 | 400V | 300.0V | 400Hz | 296.5 to 303.5 | 296.5 to 303.5 |
|
|
|
|
|
|
|
|
|
| 9 | 750V | 750V | 60Hz | 736 to 764 | 736 to 764 |
|
|
|
|
|
|
|
|
|
| 10 | 750V | 750V | 400Hz | 736 to 764 | 736 to 764 |
|
|
|
|
|
|
|
|
|
Testing the DC microamperes Function (for 61-340 and 61-342)
To verify the accuracy of AC and DC current measurement functions, do the following:
1.Connect the calibrator to the and COM inputs on the meter.
2.Turn the rotary switch to
a. Press the SELECT button to select the DC function
3.Apply the inputs for steps
4.For each input, compare the reading on the meter display to the reading for your meter in Table 3a
5.If the display reading falls outside of the range shown in Table 3a, the meter does not meet specification.
Table 3a DC microamperes Test:
Step | Range | Source |
1 | 400µA | 4.0µA |
|
|
|
2 | 400µA | 300.0µA |
|
|
|
3 | 4000µA | 3000µA |
|
|
|
Reading
3.5to 4.5
297.4to 302.6
2974 to 3026
Testing the DC milli amperes Function (for 61-340 and 61-342)
1.Connect the calibrator to the and COM inputs on the meter.
2.Turn the rotary switch to .
a. Press the SEL button to select the DC function
3.Apply the inputs for steps
4.For each input, compare the reading on the meter display to the reading for your meter in Table 3b
5.If the display reading falls outside of the range shown in Table 3b, the meter does not meet specification.
Form Number | Rev 2 November 2007 |