Table 2 AC Voltage Test:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Step

Range

Input

Frequency

Reading 61-340

Reading 61-342

 

 

 

 

 

 

 

 

 

 

1

400mV

300mV

60

295.6 to304.4

295.6 to304.4

 

 

 

 

 

 

 

 

 

 

2

400mV

300mV

400

295.6 to304.4

295.6 to304.4

 

 

 

 

 

 

 

 

 

 

3

4V

3.000V

60Hz

2.965 to 3.035

2.965 to 3.035

 

 

 

 

 

 

 

 

 

 

4

4V

3.000V

400Hz

2.965 to 3.035

2.965 to 3.035

 

 

 

 

 

 

 

 

 

 

5

40V

30.00V

60Hz

29.65 to 30.35

29.65 to 30.35

 

 

 

 

 

 

 

 

 

 

6

40V

30.00V

400Hz

29.65 to 30.35

29.65 to 30.35

 

 

 

 

 

 

 

 

 

 

7

400V

300.0V

60Hz

296.5 to 303.5

296.5 to 303.5

 

 

 

 

 

 

 

 

 

 

8

400V

300.0V

400Hz

296.5 to 303.5

296.5 to 303.5

 

 

 

 

 

 

 

 

 

 

9

750V

750V

60Hz

736 to 764

736 to 764

 

 

 

 

 

 

 

 

 

 

10

750V

750V

400Hz

736 to 764

736 to 764

 

 

 

 

 

 

 

 

 

Testing the DC microamperes Function (for 61-340 and 61-342)

To verify the accuracy of AC and DC current measurement functions, do the following:

1.Connect the calibrator to the and COM inputs on the meter.

2.Turn the rotary switch to

a. Press the SELECT button to select the DC function

3.Apply the inputs for steps 1-3 in Table 3a.

4.For each input, compare the reading on the meter display to the reading for your meter in Table 3a

5.If the display reading falls outside of the range shown in Table 3a, the meter does not meet specification.

Table 3a DC microamperes Test:

Step

Range

Source

1

400µA

4.0µA

 

 

 

2

400µA

300.0µA

 

 

 

3

4000µA

3000µA

 

 

 

Reading

3.5to 4.5

297.4to 302.6

2974 to 3026

Testing the DC milli amperes Function (for 61-340 and 61-342)

1.Connect the calibrator to the and COM inputs on the meter.

2.Turn the rotary switch to .

a. Press the SEL button to select the DC function

3.Apply the inputs for steps 1-4 in Table 3b.

4.For each input, compare the reading on the meter display to the reading for your meter in Table 3b

5.If the display reading falls outside of the range shown in Table 3b, the meter does not meet specification.