Testing the Voltage Function (for 61-340 / 61-342)
To verify accuracy in the AC and DC voltage ranges, do the following:
1. Turn the rotary switch to “ ” position.
2. Connect the calibrator to the | and COM inputs on the meter. |
3.Set the calibrator for the voltage from step 1 to 7 in Table 1.
4.Compare the reading on the meter display with the display reading shown in Table 1.
5.If the display reading falls outside of the range shown in Table 1, the meter does not meet specification.
Table 1 DC Voltage Test:
| Step | Range |
| Input | Reading |
| 1 | 400mV | |||
|
|
|
|
|
|
| 2 | 400mV |
| 4.0mV | 3.6 to 4.4 |
|
|
|
|
|
|
| 3 | 400mV |
| 300.0mV | 297.3 to 302.7 |
|
|
|
|
|
|
| 4 | 4V |
| 3.000V | 2.975 to 3.025 |
|
|
|
|
|
|
| 5 | 40V |
| 30.00V | 29.75 to 30.25 |
|
|
|
|
|
|
| 6 | 400V |
| 300.0V | 297.5 to 302.5 |
|
|
|
|
|
|
| 7 | 1000V |
| 1000V | 987 to 1013 |
|
|
|
|
|
|
1. Turn the rotary switch to position. “ | ” |
|
2.Set the calibration for the voltage and frequency from step 1 to 10 in Table 2.
3.Compare the reading on the meter display with the display reading shown in Table 2.
4.If the display reading falls outside of the range shown in Table 2, the meter does not meet specification.