Page 14

Testing the Voltage Function (for 61-340 / 61-342)

To verify accuracy in the AC and DC voltage ranges, do the following:

1. Turn the rotary switch to “ ” position.

2. Connect the calibrator to the

and COM inputs on the meter.

3.Set the calibrator for the voltage from step 1 to 7 in Table 1.

4.Compare the reading on the meter display with the display reading shown in Table 1.

5.If the display reading falls outside of the range shown in Table 1, the meter does not meet specification.

Table 1 DC Voltage Test:

 

Step

Range

 

Input

Reading

 

1

400mV

-300.0mV

-297.3 to -302.7

 

 

 

 

 

 

 

2

400mV

 

4.0mV

3.6 to 4.4

 

 

 

 

 

 

 

3

400mV

 

300.0mV

297.3 to 302.7

 

 

 

 

 

 

 

4

4V

 

3.000V

2.975 to 3.025

 

 

 

 

 

 

 

5

40V

 

30.00V

29.75 to 30.25

 

 

 

 

 

 

 

6

400V

 

300.0V

297.5 to 302.5

 

 

 

 

 

 

 

7

1000V

 

1000V

987 to 1013

 

 

 

 

 

 

1. Turn the rotary switch to position.

 

2.Set the calibration for the voltage and frequency from step 1 to 10 in Table 2.

3.Compare the reading on the meter display with the display reading shown in Table 2.

4.If the display reading falls outside of the range shown in Table 2, the meter does not meet specification.

Form Number TM61340-2

Rev 2 November 2007

Page 16
Image 16
IDEAL INDUSTRIES technical manual Testing the Voltage Function for 61-340, Step Range Input Reading