Page 23

6.Using a small flat-tipped screwdriver, adjust VRtr until the display reads 2.995 to 3.005Volts

7.Disconnect the AC calibrator from the meter.

(C)TEMP Calibration (Adjust VR3)

1.Set the rotary switch to the " ºF/ ºC " position.

2.Set the output of the calibrator to 300 ºC

3.Connect the calibrator temperature output to the and COM input terminals

4.Using a small flat-tipped screwdriver to adjust VR3 for a 300 ºC +/- 1 ºC (299 to 301) display

Form Number TM61340-2

Rev 2 November 2007