Page 23
6.Using a small
7.Disconnect the AC calibrator from the meter.
(C)TEMP Calibration (Adjust VR3)
1.Set the rotary switch to the " ºF/ ºC " position.
2.Set the output of the calibrator to 300 ºC
3.Connect the calibrator temperature output to the and COM input terminals
4.Using a small
Form Number | Rev 2 November 2007 |