Page 17

Testing the AC milliampere function: (for 61-340and 61-342)

1.Connect the calibrator to the and COM inputs on the meter.

2.Turn the rotary switch to .

a.Press the SELECT button to select the AC function

b.Press the REL button to make the display is zero

3.Apply the inputs for steps 1-4 in Table 4b.

4.For each input, compare the reading on the meter display to the reading for your meter in Table 4b

5.If the display reading falls outside of the range shown in Table 4b, the meter does not meet specification.

Table 4b ACmA Test:

Step

Range

Source

Frequency

Reading

1

40 mA

30.00mA

60Hz

29.35 to 30.65

 

 

 

 

 

2

40 mA

30.00mA

400Hz

29.35 to 30.65

 

 

 

 

 

3

400 mA

300.0mA

60Hz

293.5 to 306.5

 

 

 

 

 

4

400 mA

300.0mA

400Hz

293.5 to 306.5

 

 

 

 

 

Testing the AC A function (for 61-340 / 61-342)

1.Connect the calibrator to the 10A and COM inputs on the meter.

2.Turn the rotary switch to .

a.Press the SELECT button to select the AC function

b.Press the REL button to make the display is zero

3.Apply the inputs for steps 1-4 in Table 4c.

4.For each input, compare the reading on the meter display to the reading for your meter in Table 4c

5.If the display reading falls outside of the range shown in Table 4c, the meter does not meet specification.

Table 4c AC A Test:

Step

 

Source

Frequency

Reading

1

4A

3.000A

60Hz

2.920 to 3.080

 

 

 

 

 

2

4A

3.000A

400Hz

2.920 to 3.080

 

 

 

 

 

3

10A

10A

60Hz

9.70 to 10.30

 

 

 

 

 

4

10A

10A

400Hz

9.70 to 10.30

 

 

 

 

 