Page 5

(3b) DC Current (61-340 / 61-342)

Range

Resolution

Accuracy

Input Protection

4.000A

0.001A

±(1.5% reading + 5 digits)

10A, 250V Fast Blow

10.00A*

0.01A

Fuse

 

Overload Protection: A Input: 10A, 250V Fast Blow fuse. (61-340 / 61-342)

*Caution: Do not make high current measurements on the 10A scale for longer that 15 seconds. This should be followed by a 15 minute cool down period. Exceeding 15 seconds may cause damage to the meter and/or the test leads.

(4a) AC micro-amp and milli-amps Current (61-340 / 61-342)

Range

Resolution

Accuracy

400.0µA

0.1μA

±(1.5% reading + 5 digits) *1

 

 

40Hz ~ 400Hz

4000µA

1μA

 

40.00mA

10μA

±(2.0% reading + 5 digits) *1

 

 

40Hz ~ 400Hz

400.0mA

0.1mA

 

Input Protection

500mA, 250V Fast

Blow Fuse

Overload Protection: μA / mA Input: 500mA,250V Fast Blow fuse. (61-340 / 61-342)

(4b) AC Current (61-340 / 61-342)

Range

Resolution

Accuracy

Input Protection

4.000A

0.001mA

±(2.5% reading + 5 digits) *1

10A, 250V Fast Blow

10.00A*

0.01mA

40Hz ~ 400Hz

Fuse.

Overload Protection: A Input: 10A (250V) fast blow fuse.

AC Conversion Type: AC conversions are True RMS responding, calibrated to the sine wave input. *1 The specified accuracy is for sine wave at full scale and non-sine wave at half scale with crest factor

up to 2.

*Caution: Do not make high current measurements on the 10A scale for longer that 15 seconds. This should be followed by a 15 minute cool down period. Exceeding 15 seconds may cause damage to the meter and/or the test leads.

Form Number TM61340-2

Rev 2 November 2007

Page 7
Image 7
IDEAL INDUSTRIES 61-340 technical manual Range Resolution Accuracy