Setting up Polarized Light (cont.)

Setting up Polarized Light (cont.)

the Analyzer into position. The Analyzer is located in the module Manual background between the viewing tube and the stand either in a slider module or in an Analyzer Bertrand Lens module. If you have a slider, slide

the Analyzer into position.

3.If you have an Analyzer/Bertrand Lens module, swing in the Analyzer by flipping the lower rocker of the module to the right (the "A" position/Analyzer in), while ensuring the upper rocker of the Analyzer/ Bertrand Lens Module is flipped to the left (the "O" position/Bertrand Lens out).

4.Increase the illumination intensity slightly. You will notice while looking into the microscope with the Polarizer and Analyzer

in place and no specimen, that the field will look dark. If not, rotate the polarizer slightly until you have the darkest field of view.

Leica DM750 P