Time Delay Shutoff
The Leica DM 750 P is equipped with a time delay shutoff capability, which automatically turns the illumination off after 2
hours of no changes in the intensity control.
ϘϘ The time delay shutoff IS enabled as a default.
Changing the status of the time delay shutoff
1.Rotate the intensity control to the lowest setting.
2.Turn the power on.
3.Rotate the intensity control to full high intensity, then back to full low intensity all within 5 seconds.
ϘϘ The LED Illumination will flash to indicate the time delay shutoff status was changed.
ϘϘ The LED Illumination will flash two times slowly then stay on when you disable the Time Delay Shutoff.
ϘϘ The LED Illumination will flash three times quickly then stay on when you enable the Time Delay Shutoff.
When you turn the power off and then turn the power on again, the system will be in the last Time Delay Shutoff status (Enabled
or Disabled) and you will not see any blinking.
Leica DM750 P | Manual | 35 |
|
|
|