Meridian Audio G61 manual Calibrating the system, Introduction, Using a Sound Pressure Level meter

Models: G61

1 59
Download 59 pages 26.08 Kb
Page 30
Image 30
Calibrating the system

2 5

C a l i b r a t i n g t h e s y s t e m

Calibrating the system

To help you to set up the installation to give the best possible sound with any particular

combination of associated equipment, the G61 Digital Surround Controller includes a built-in

calibration procedure.

This calibration procedure uses test signals to present a series of sounds, which you use to

adjust certain aspects of the system to their optimum settings.

You should use the calibration procedure the first time you set up the system, and whenever

you want to check the calibration, such as after changing the layout of the room.

Introduction

We recommend that you perform the calibration procedure using the MSR+ and from the listening position.

As you run the calibration procedure the name of each calibration test is shown on the front-panel display, followed by the parameters adjusted in the test.

Each calibration test uses a test signal designed to give the best results.

Using a Sound Pressure Level meter

Although you can perform the calibration procedure by ear, it is recommended that you perform the tests using a Sound Pressure Level meter. These are available fairly cheaply from Tandy/Radio Shack, or your Meridian dealer may be able to lend you one.

Set the Sound Pressure Level meter to C weighted, and Slow. Take readings with the meter at the listening position, pointing vertically. You should hold the meter with an outstretched arm to minimise reflections from your body.

You can choose to display and enter distances in either feet and inches or metres.

Page 30
Image 30
Meridian Audio G61 manual Calibrating the system, Introduction, Using a Sound Pressure Level meter