E1 - Reverse Elevation Pot

Step 1. Calibrate Elevation

Calibrate the elevation as outlined in the System Calibration Procedure in below step.

Error 1

Enter engineer mode of

“MANUAL

CALIBRATION”

screen

Check the “P13 ELV DIRECTION” screen whether set to “Yes”

Enter the “P4 ELEVATION

MINIMUM” screen

whether parameter set is

30 2

Enter the “P5

ELEVATION

MAXIMUM ” screen

whether parameter set is

240 5

Set to “Yes”

AUTO-

Calibration

 

AUTO-

Calibration

AUTO-

Calibration

Step 2. CORRECTIVE ACTION

1.Ensure correct signal is being sent to interface board (MCB) by pressing elevation up/down keys by observing elevation LED’s correctly corresponding to correct elevation keys pressed (please refer the 29 page) . If failure occurs for this step check console cable and then the interface board (MCB).

2.Check failed elevation motor, interface board, console cable with a known good console.

32

Page 32
Image 32