E2 – Elevation Out Of Range

Step 1. Calibrate Elevation

Calibrate the elevation as outlined in the System Calibration Procedure in below step.

Error 2

Enter engineer mode of

“MANUAL

CALIBRATION”

screen

Check the “P13 ELV DIRECTION” screen whether set to “Yes”

Enter the “P4 ELEVATION

MINIMUM” screen

whether parameter set is

30 2

Enter the “P5

ELEVATION

MAXIMUM ” screen

whether parameter set is

240 5

Step 2. CORRECTIVE ACTION

Set to “Yes”

AUTO-

Calibration

 

AUTO-

Calibration

AUTO-

Calibration

1.Check the failed system with a known good console cable. If fixed then repair/replace the console cable.

2.Check the failed system with a known good elevation motor. If problem fixed then repair/replace the elevation motor.

3.Check the failed system with a known good interface board. If problem fixed then repair/replace the interface board (MCB).

4.Check the failed system with a known good console board with current software. If the problem is fixed then repair/replace the console.