Appendix C Advanced Hardware Configuration Settings
© National Instruments Corporation C-7 VXI-MXI-Express Series User Manual
TRIG/CLK PORT A(B) DirectionW7—Trigger Card CLK10 Dir
This jumper allows you to configure the direction of CLK10 through the
TRIG/CLK PORT A(B).
In Slot 0 operation with jumpers W4 set for an external source and W3
set for the front panel TRIG/CLK PORT A(B) set this jumper to In, as
shown in FigureC-7a, to route CLK10 in from another NI VXI-8360T
to the VXI backplane. In non-slot0 operation the In setting has no
effect.
In Slot 0 and non-Slot 0 operation setting this jumper to Out,
as shown in FigureC-7b, routes the VXIbus CLK10 out of the
TRIG/CLKPORT A(B).
Figure C-7. Setting the TRIG/CLK PORT A(B) Direction with Jumper W7
Optional CLK10 SettingsInverting CLK10
If the front panel SMB connector EXT CLK is configured as an output this
software setting provides the capability to invert the clock output. This
setting can be found in the hardware configuration panels in MAX as
shown in FigureC-8.
Set this jumper to OUT to route CLK10
out of the TRIG/CLK PORT.
Set this jumper to IN when the NI VXI-8360T
is in Slot 0 and W3 and W4 are configured to
select the TRIG/CLK PORT as the external
CLK10 source. Otherwise leave W7 set to IN
for other W3 and W4 combinations.