OMB-DAQ-2416-4AO User's Guide

 

 

 

Specifications

 

 

 

Table 5. K type thermocouple accuracy specifications, including CJC measurement error.

 

 

 

 

 

All specifications are (±).

 

 

 

 

 

 

 

 

 

 

 

 

 

A/D data

 

Sensor temperature

 

Accuracy error

Accuracy error

Tempco

 

 

rate

 

range

 

maximum

typical

(°C/°C)

 

 

 

 

 

 

 

 

 

 

 

3750 S/s

 

-210 °C

 

2.917 °C

1.699 °C

0.029

 

 

 

 

0 °C

 

1.017 °C

0.526 °C

 

 

 

 

 

1372 °C

 

2.478 °C

2.022 °C

 

 

 

2000 S/s

 

-210 °C

 

2.917 °C

1.699 °C

0.029

 

 

 

 

 

 

 

 

 

 

 

 

 

0 °C

 

1.017 °C

0.526 °C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1372 °C

 

2.478 °C

2.022 °C

 

 

 

 

 

 

 

 

 

 

 

 

1000 S/s

 

-210 °C

 

2.917 °C

1.699 °C

0.029

 

 

 

 

 

 

 

 

 

 

 

 

 

0 °C

 

1.017 °C

0.526 °C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1372 °C

 

2.478 °C

2.022 °C

 

 

 

 

 

 

 

 

 

 

 

 

500 S/s

 

-210 °C

 

2.735 °C

1.524 °C

0.029

 

 

 

 

0 °C

 

0.948 °C

0.457 °C

 

 

 

 

 

1372 °C

 

2.396 °C

1.941 °C

 

 

 

 

 

 

 

 

 

 

 

 

100 S/s

 

-210 °C

 

2.735 °C

1.524 °C

0.029

 

 

 

 

 

 

 

 

 

 

 

 

 

0 °C

 

0.948 °C

0.457 °C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1372 °C

 

2.396 °C

1.941 °C

 

 

 

 

 

 

 

 

 

 

 

 

60 S/s

 

-210 °C

 

2.735 °C

1.524 °C

0.029

 

 

 

 

0 °C

 

0.948 °C

0.457 °C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1372 °C

 

2.396 °C

1.941 °C

 

 

 

50 S/s

 

-210 °C

 

2.735 °C

1.524 °C

0.029

 

 

 

 

 

 

 

 

 

 

 

 

 

0 °C

 

0.948 °C

0.457 °C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1372 °C

 

2.396 °C

1.941 °C

 

 

 

 

 

 

 

 

 

 

 

 

25 S/s

 

-210 °C

 

2.735 °C

1.524 °C

0.029

 

 

 

 

 

 

 

 

 

 

 

 

 

0 °C

 

0.948 °C

0.457 °C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1372 °C

 

2.396 °C

1.941 °C

 

 

 

 

 

 

 

 

 

 

 

 

10 S/s

 

-210 °C

 

2.735 °C

1.524 °C

0.029

 

 

 

 

0 °C

 

0.948 °C

0.457 °C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1372 °C

 

2.396 °C

1.941 °C

 

 

 

 

 

 

 

 

 

 

 

 

5 S/s

 

-210 °C

 

2.735 °C

1.524 °C

0.029

 

 

 

 

 

 

 

 

 

 

 

 

 

0 °C

 

0.948 °C

0.457 °C

 

 

 

 

 

 

 

 

 

 

 

 

 

 

1372 °C

 

2.396 °C

1.941 °C

 

 

 

 

 

 

 

 

 

 

 

 

2.5 S/s

 

-210 °C

 

2.735 °C

1.524 °C

0.029

 

 

 

 

0 °C

 

0.948 °C

0.457 °C

 

 

 

 

 

1372 °C

 

2.396 °C

1.941 °C

 

 

To achieve the thermocouple accuracies listed above, the OMB-DAQ-2416-4AO should be warmed up for

45 minutes after the initial power on. The accuracies listed above are only guaranteed if the OMB-DAQ-2416- 4AO is housed in its plastic enclosure.

28