OMB-DAQ-2416-4AO User's Guide

 

 

Specifications

 

Table 7. R type thermocouple accuracy specifications, including CJC measurement error.

 

 

 

 

 

All specifications are (±).

 

 

 

 

 

 

 

 

 

 

 

A/D data rate

Sensor

Accuracy error

Accuracy error

 

Tempco

 

 

temperature range

maximum

typical

 

(°C/°C)

 

 

 

 

 

 

 

 

 

 

 

 

3750 S/s

-50 °C

4.826 °C

3.133 °C

 

0.082

 

 

 

 

 

 

 

 

 

250 °C

2.117 °C

1.424 °C

 

 

 

 

 

 

 

 

 

 

 

1768 °C

2.842 °C

2.347 °C

 

 

 

2000 S/s

-50 °C

4.826 °C

3.133 °C

 

0.082

 

 

250 °C

2.117 °C

1.424 °C

 

 

 

 

 

 

 

 

 

 

 

1768 °C

2.842 °C

2.347 °C

 

 

 

 

 

 

 

 

 

 

1000 S/s

-50 °C

4.826 °C

3.133 °C

 

0.082

 

 

 

 

 

 

 

 

 

250 °C

2.117 °C

1.424 °C

 

 

 

 

 

 

 

 

 

 

 

1768 °C

2.842 °C

2.347 °C

 

 

 

 

 

 

 

 

 

 

500 S/s

-50 °C

4.065 °C

2.379 °C

 

0.082

 

 

250 °C

1.805 °C

1.113 °C

 

 

 

 

1768 °C

2.619 °C

2.123 °C

 

 

 

100 S/s

-50 °C

4.065 °C

2.379 °C

 

0.082

 

 

 

 

 

 

 

 

 

250 °C

1.805 °C

1.113 °C

 

 

 

 

 

 

 

 

 

 

 

1768 °C

2.619 °C

2.123 °C

 

 

 

 

 

 

 

 

 

 

60 S/s

-50 °C

4.065 °C

2.379 °C

 

0.082

 

 

 

 

 

 

 

 

 

250 °C

1.805 °C

1.113 °C

 

 

 

 

 

 

 

 

 

 

 

1768 °C

2.619 °C

2.123 °C

 

 

 

 

 

 

 

 

 

 

50 S/s

-50 °C

4.065 °C

2.379 °C

 

0.082

 

 

250 °C

1.805 °C

1.113 °C

 

 

 

 

 

 

 

 

 

 

 

1768 °C

2.619 °C

2.123 °C

 

 

 

 

 

 

 

 

 

 

25 S/s

-50 °C

4.065 °C

2.379 °C

 

0.082

 

 

 

 

 

 

 

 

 

250 °C

1.805 °C

1.113 °C

 

 

 

 

 

 

 

 

 

 

 

1768 °C

2.619 °C

2.123 °C

 

 

 

 

 

 

 

 

 

 

10 S/s

-50 °C

4.065 °C

2.379 °C

 

0.082

 

 

250 °C

1.805 °C

1.113 °C

 

 

 

 

1768 °C

2.619 °C

2.123 °C

 

 

 

 

 

 

 

 

 

 

5 S/s

-50 °C

4.065 °C

2.379 °C

 

0.082

 

 

 

 

 

 

 

 

 

250 °C

1.805 °C

1.113 °C

 

 

 

 

 

 

 

 

 

 

 

1768 °C

2.619 °C

2.123 °C

 

 

 

 

 

 

 

 

 

 

2.5 S/s

-50 °C

4.065 °C

2.379 °C

 

0.082

 

 

250 °C

1.805 °C

1.113 °C

 

 

 

 

 

 

 

 

 

 

 

1768 °C

2.619 °C

2.123 °C

 

 

To achieve the thermocouple accuracies listed above, the OMB-DAQ-2416-4AO should be warmed up for

45 minutes after the initial power on. The accuracies listed above are only guaranteed if the OMB-DAQ-2416- 4AO is housed in its plastic enclosure.

30