OMB-DAQ-2416-4AO User's Guide

 

 

Specifications

 

 

Table 6. N type thermocouple accuracy specifications, including CJC measurement error.

 

 

 

 

All specifications are (±).

 

 

 

 

 

 

 

 

 

 

A/D data

 

Sensor

Accuracy error maximum

Accuracy error

Tempco

 

rate

 

temperature range

 

typical

(°C/°C)

 

 

 

 

 

 

 

 

3750 S/s

 

-200 °C

3.480°C

2.030 °C

0.029

 

 

 

 

 

 

 

 

 

 

0 °C

1.201 °C

0.659 °C

 

 

 

 

 

 

 

 

 

 

 

1300 °C

1.991 °C

1.600 °C

 

 

2000 S/s

 

-200 °C

3.480°C

2.030 °C

0.029

 

 

 

0 °C

1.201 °C

0.659 °C

 

 

 

 

 

 

 

 

 

 

 

1300 °C

1.991 °C

1.600 °C

 

 

 

 

 

 

 

 

 

1000 S/s

 

-200 °C

3.480°C

2.030 °C

0.029

 

 

 

 

 

 

 

 

 

 

0 °C

1.201 °C

0.659 °C

 

 

 

 

 

 

 

 

 

 

 

1300 °C

1.991 °C

1.600 °C

 

 

 

 

 

 

 

 

 

500 S/s

 

-200 °C

3.196°C

1.750 °C

0.029

 

 

 

0 °C

1.096 °C

0.553 °C

 

 

 

 

1300 °C

1.915 °C

1.524 °C

 

 

100 S/s

 

-200 °C

3.196°C

1.750 °C

0.029

 

 

 

 

 

 

 

 

 

 

0 °C

1.096 °C

0.553 °C

 

 

 

 

 

 

 

 

 

 

 

1300 °C

1.915 °C

1.524 °C

 

 

 

 

 

 

 

 

 

60 S/s

 

-200 °C

3.196°C

1.750 °C

0.029

 

 

 

 

 

 

 

 

 

 

0 °C

1.096 °C

0.553 °C

 

 

 

 

 

 

 

 

 

 

 

1300 °C

1.915 °C

1.524 °C

 

 

 

 

 

 

 

 

 

50 S/s

 

-200 °C

3.196°C

1.750 °C

0.029

 

 

 

0 °C

1.096 °C

0.553 °C

 

 

 

 

 

 

 

 

 

 

 

1300 °C

1.915 °C

1.524 °C

 

 

 

 

 

 

 

 

 

25 S/s

 

-200 °C

3.196°C

1.750 °C

0.029

 

 

 

 

 

 

 

 

 

 

0 °C

1.096 °C

0.553 °C

 

 

 

 

 

 

 

 

 

 

 

1300 °C

1.915 °C

1.524 °C

 

 

 

 

 

 

 

 

 

10 S/s

 

-200 °C

3.196°C

1.750 °C

0.029

 

 

 

0 °C

1.096 °C

0.553 °C

 

 

 

 

1300 °C

1.915 °C

1.524 °C

 

 

 

 

 

 

 

 

 

5 S/s

 

-200 °C

3.196°C

1.750 °C

0.029

 

 

 

 

 

 

 

 

 

 

0 °C

1.096 °C

0.553 °C

 

 

 

 

 

 

 

 

 

 

 

1300 °C

1.915 °C

1.524 °C

 

 

 

 

 

 

 

 

 

2.5 S/s

 

-200 °C

3.196°C

1.750 °C

0.029

 

 

 

0 °C

1.096 °C

0.553 °C

 

 

 

 

 

 

 

 

 

 

 

1300 °C

1.915 °C

1.524 °C

 

To achieve the thermocouple accuracies listed above, the OMB-DAQ-2416-4AO should be warmed up for

45 minutes after the initial power on. The accuracies listed above are only guaranteed if the OMB-DAQ-2416- 4AO is housed in its plastic enclosure.

29