OMB-DAQ-2416-4AO User's Guide

 

 

Specifications

 

Table 10. E type thermocouple accuracy specifications, including CJC measurement error.

 

 

 

All specifications are (±).

 

 

 

 

 

 

 

 

 

A/D data

Sensor temperature

Accuracy error

Accuracy error typical

Tempco

 

rate

range

maximum

 

(°C/°C)

 

 

 

 

 

 

 

3750 S/s

-200 °C

2.413 °C

1.352 °C

0.017

 

 

 

 

 

 

 

 

0 °C

1.069 °C

0.551 °C

 

 

 

 

 

 

 

 

 

1000 °C

1.575 °C

1.211 °C

 

 

 

 

 

 

 

 

2000 S/s

-200 °C

2.413 °C

1.352 °C

0.017

 

 

0 °C

1.069 °C

0.551 °C

 

 

 

 

 

 

 

 

 

1000 °C

1.575 °C

1.211 °C

 

 

 

 

 

 

 

 

1000 S/s

-200 °C

2.413 °C

1.352 °C

0.017

 

 

 

 

 

 

 

 

0 °C

1.069 °C

0.551 °C

 

 

 

 

 

 

 

 

 

1000 °C

1.575 °C

1.211 °C

 

 

 

 

 

 

 

 

500 S/s

-200 °C

2.306 °C

1.244 °C

0.017

 

 

0 °C

1.017 °C

0.499 °C

 

 

 

1000 °C

1.539 °C

1.175 °C

 

 

100 S/s

-200 °C

2.306 °C

1.244 °C

0.017

 

 

 

 

 

 

 

 

0 °C

1.017 °C

0.499 °C

 

 

 

 

 

 

 

 

 

1000 °C

1.539 °C

1.175 °C

 

 

 

 

 

 

 

 

60 S/s

-200 °C

2.306 °C

1.244 °C

0.017

 

 

 

 

 

 

 

 

0 °C

1.017 °C

0.499 °C

 

 

 

 

 

 

 

 

 

1000 °C

1.539 °C

1.175 °C

 

 

 

 

 

 

 

 

50 S/s

-200 °C

2.306 °C

1.244 °C

0.017

 

 

0 °C

1.017 °C

0.499 °C

 

 

 

1000 °C

1.539 °C

1.175 °C

 

 

 

 

 

 

 

 

25 S/s

-200 °C

2.306 °C

1.244 °C

0.017

 

 

 

 

 

 

 

 

0 °C

1.017 °C

0.499 °C

 

 

 

 

 

 

 

 

 

1000 °C

1.539 °C

1.175 °C

 

 

 

 

 

 

 

 

10 S/s

-200 °C

2.306 °C

1.244 °C

0.017

 

 

0 °C

1.017 °C

0.499 °C

 

 

 

1000 °C

1.539 °C

1.175 °C

 

 

5 S/s

-200 °C

2.306 °C

1.244 °C

0.017

 

 

 

 

 

 

 

 

0 °C

1.017 °C

0.499 °C

 

 

 

 

 

 

 

 

 

1000 °C

1.539 °C

1.175 °C

 

 

 

 

 

 

 

 

2.5 S/s

-200 °C

2.306 °C

1.244 °C

0.017

 

 

 

 

 

 

 

 

0 °C

1.017 °C

0.499 °C

 

 

 

 

 

 

 

 

 

1000 °C

1.539 °C

1.175 °C

 

To achieve the thermocouple accuracies listed above, the OMB-DAQ-2416-4AO should be warmed up for

45 minutes after the initial power on. The accuracies listed above are only guaranteed if the OMB-DAQ-2416- 4AO is housed in its plastic enclosure.

33