OMB-DAQ-2416-4AO User's Guide

 

 

Specifications

 

Table 11. T type thermocouple accuracy specifications, including CJC measurement error.

 

 

 

 

 

All specifications are (±).

 

 

 

 

 

 

 

 

 

 

 

A/D data rate

Sensor

Accuracy error

Accuracy error

 

Tempco

 

 

temperature range

maximum

typical

 

(°C/°C)

 

 

 

 

 

 

 

 

 

 

 

 

3750 S/s

-200 °C

2.821 °C

1.676 °C

 

0.027

 

 

 

 

 

 

 

 

 

0 °C

1.050 °C

0.558 °C

 

 

 

 

 

 

 

 

 

 

 

400 °C

0.957 °C

0.595 °C

 

 

 

 

 

 

 

 

 

 

2000 S/s

-200 °C

2.821 °C

1.676 °C

 

0.027

 

 

0 °C

1.050 °C

0.558 °C

 

 

 

 

 

 

 

 

 

 

 

400 °C

0.957 °C

0.595 °C

 

 

 

 

 

 

 

 

 

 

1000 S/s

-200 °C

2.821 °C

1.676 °C

 

0.027

 

 

 

 

 

 

 

 

 

0 °C

1.050 °C

0.558 °C

 

 

 

 

 

 

 

 

 

 

 

400 °C

0.957 °C

0.595 °C

 

 

 

 

 

 

 

 

 

 

500 S/s

-200 °C

2.644 °C

1.505 °C

 

0.027

 

 

0 °C

0.979 °C

0.487 °C

 

 

 

 

400 °C

0.912 °C

0.551 °C

 

 

 

100 S/s

-200 °C

2.644 °C

1.505 °C

 

0.027

 

 

 

 

 

 

 

 

 

0 °C

0.979 °C

0.487 °C

 

 

 

 

 

 

 

 

 

 

 

400 °C

0.912 °C

0.551 °C

 

 

 

 

 

 

 

 

 

 

60 S/s

-200 °C

2.644 °C

1.505 °C

 

0.027

 

 

 

 

 

 

 

 

 

0 °C

0.979 °C

0.487 °C

 

 

 

 

 

 

 

 

 

 

 

400 °C

0.912 °C

0.551 °C

 

 

 

 

 

 

 

 

 

 

50 S/s

-200 °C

2.644 °C

1.505 °C

 

0.027

 

 

0 °C

0.979 °C

0.487 °C

 

 

 

 

400 °C

0.912 °C

0.551 °C

 

 

 

 

 

 

 

 

 

 

25 S/s

-200 °C

2.644 °C

1.505 °C

 

0.027

 

 

 

 

 

 

 

 

 

0 °C

0.979 °C

0.487 °C

 

 

 

 

 

 

 

 

 

 

 

400 °C

0.912 °C

0.551 °C

 

 

 

 

 

 

 

 

 

 

10 S/s

-200 °C

2.644 °C

1.505 °C

 

0.027

 

 

0 °C

0.979 °C

0.487 °C

 

 

 

 

400 °C

0.912 °C

0.551 °C

 

 

 

5 S/s

-200 °C

2.644 °C

1.505 °C

 

0.027

 

 

 

 

 

 

 

 

 

0 °C

0.979 °C

0.487 °C

 

 

 

 

 

 

 

 

 

 

 

400 °C

0.912 °C

0.551 °C

 

 

 

 

 

 

 

 

 

 

2.5 S/s

-200 °C

2.644 °C

1.505 °C

 

0.027

 

 

 

 

 

 

 

 

 

0 °C

0.979 °C

0.487 °C

 

 

 

 

 

 

 

 

 

 

 

400 °C

0.912 °C

0.551 °C

 

 

To achieve the thermocouple accuracies listed above, the OMB-DAQ-2416-4AO should be warmed up for

45 minutes after the initial power on. The accuracies listed above are only guaranteed if the OMB-DAQ-2416- 4AO is housed in its plastic enclosure.

34