Omega Speaker Systems OS523E, OS532E, OS524E, OS530LE, OS53xE-CF If necessary, adjust the emissivity

Models: OS53xE-CF OS524E OS530HRE

1 84
Download 84 pages 50.27 Kb
Page 22
Image 22
2.If necessary, adjust the emissivity.

Using the Handheld Infrared Thermometer 2

Static Surface Scan – Measures the temperature across a static surface:

1.Aim the thermometer at a starting point and pull the trigger. Press the LOCK key to lock the trigger.

2.If necessary, adjust the emissivity.

3.Slowly move the thermometer so that the line of sight sweeps across the surface. The thermometer measures the temperature at each point on the surface.

4.To record the temperature profile across the surface, connect the IR thermometer to a strip chart recorder. Refer to Figure

2-11 for details. The IR thermometer provides an analog output of 1mV/degree. (0.5 mV/Deg on OS524E)

5.After all the data has been taken, press the LOCK key to unlock the trigger.

NOTE

Center hole is the

analog output jack

Analog

Cable

To Strip Chart

Recorder

Figure 2-11 Recorder Hookup

Moving Surface Scan - Measures the Temperature of Points on a Moving Surface:

1.Mount the thermometer on a camera tripod and aim at a fixed point on the moving surface.

2.Pull the trigger and press the LOCK key to lock the trigger.

3.If necessary, adjust the emissivity. The thermometer is now set up for measuring the temperature of a moving surface.

4.To record the temperature profile of the moving surface, connect the IR thermometer to a strip chart recorder. Refer to Figure 2-11 for details.

5.After all data is taken, press the LOCK key to unlock the trigger.

Fixed Point Monitoring Over Time - Monitors the temperature at a fixed point over time:

NOTE

It is recommended that you use the ac adapter for long term measurement of temperature.

2-7

Page 22
Image 22
Omega Speaker Systems OS523E, OS532E, OS524E, OS530LE, OS53xE-CF, OS534E, OS530HRE, OS533E If necessary, adjust the emissivity