TROUBLE SHOOTING
If the meter locks on a particular display, unplug and plug in again to unlock the display. A total reset clears the memory and defaults all values to zero. To perform this reset,
Factory Channel settings:
To obtain factory preset values simply
Factory default settings for Thermocouple, voltage, current and
PARAMETER | CURRENT | VOLTAGE. | MILLIVOLT | T/C |
INPUT | INPUT | INPUT | INPUT | |
HIGH SCALE | 20.000 | 10.000 | 100.00 | J |
LOW SCALE | 0.000 | 0.000 | 0.000 | J |
OFFSET | 0.00 | 0.00 | 0.00 | J |
TARE | 0.000 | 0.000 | 0.000 | J |
ERROR MESSAGES AND SOLUTIONS
ERROR MESSAGES AND SOLUTIONS
ERROR MESSAGE | ERROR CONDITION | SOLUTION |
'EE EROR' | The unit was unable to | If the message keeps coming up, |
| save programmed value | unplug the unit and reapply power. |
| in EEPROM | If the unit still displays the error |
| failure of EEPROM. | message, send it in for repair. |
|
| If error message followed by |
|
| 8.8.8.8.8.8.8.and the revision |
|
| number are displayed occasionally, |
|
| it is an indication that power |
|
| line is very noisy and needs a line filter. |
'CARD ER' | Too much noise on | 1. If message shows occasionally, |
| input signals or problem | it indicates that input signal s have |
| with Scanner card. | too much high voltage noise. |
|
| Provide signal filters. |
|
| 2. Unplug and plug the meter again |
|
| to recycle power. |
|
| 3. Scanner card defective |
|
| for repair |
'HIGH AC NOISE' | High AC noise on input | 1. If message shows occasionally, |
| signals or incoming AC | it indicates that input signals or |
| line interfering with the | AC line has too much high |
| operation of the unit | voltage noise. |
|
| 2. Provide line and signal filters. |
|
|
|
Page 26