Omron Healthcare HEM-705CP manual Troubleshooting

Models: HEM-705CP

1 28
Download 28 pages 10.32 Kb
Page 21
Image 21

TROUBLESHOOTING

Main Unit:

Error Indicator

Cause

Correction

 

 

Turn monitor off, wait 2-3 minutes

 

 

and take another measurement.

 

Cuff under-inflated.

Repeat steps listed under "How

 

to Take a Reading." If "EE"

 

 

 

 

is displayed again, take another

 

 

measurement by pressing and

 

 

holding the START button

 

 

 

 

as shown on instruction page

 

Movement during

12 under "How to Take

 

a Reading", instructions

 

measurement.

for special conditions.

 

 

Your cuff maybe wrapped

 

 

too tightly around your arm.

 

 

 

 

Cuff over-inflated

Turn the monitor off, carefully

 

read and repeat steps listed

 

(more than 300 mmHg).

 

under "How to Take a Reading."

 

 

Battery voltage

is excessively low.

Replace all four "AA" batteries with new batteries.

Problem caused by abnormal memory function.

Refer to warranty page for sending unit in for repair service.

If you continue to have difficulty obtaining a reading, call

OMRON HEALTHCARE CUSTOMER SERVICE TOLL FREE 1-800-634-4350.

Please tell the representative you have an OMRON Model HEM-705CP.