Pride Mobility Jazzy 1113 owner manual Motor Brushes, To inspect or replace the motor brushes

Models: Jazzy 1113

1 60
Download 60 pages 59.9 Kb
Page 52
Image 52

X . C A R E A N D M A I N T E N A N C E

11.Place the top cover back onto the bottom cover.

12.Fasten the straps and make sure the connection is secure. Place the battery box aside.

13.For the other battery, repeat steps 4 - 12.

14.Make sure that the battery terminals are opposite the battery charger.

WARNING! If you do not position the battery terminals on the opposite side of the charger, you run the risk of shorting out the batteries and severely damaging your Jazzy!

15.Place the rear battery box back into the Jazzy battery well.

16.Place the front battery box back into the Jazzy battery well.

17.Connect the rear battery box connector and the front battery box connector to the electronics tray. See figure 36.

18.Install the seat.

19.Connect the controller cable.

MOTOR BRUSHES

The electric motors that power your Jazzy 1113 use carbon brushes. These brushes may become susceptible to wear over a long period of time. The motor brushes are the two contacts located inside the motor assembly that supply power to the motor. They are designed to provide several thousand hours of operation.

CAUTION! If the brushes become dirty with carbon deposits or wear out, a motor will run poorly or not at all. If inspection determines excessive wear on the brushes, they must be replaced or motor damage will result. Failure to maintain the brushes could void the Jazzy 1113’s warranty.

To inspect or replace the motor brushes:

1.Remove the seat and body shroud. See V. “Disassembly” in this manual.

2.Unscrew the motor brush caps. See figure 37.

3.Remove the brushes.

4.Inspect for wear. See figure 38.

5.Replace the brushes, if necessary. Contact your local Pride provider for replacement brushes.

CORRECTIVE MAINTENANCE

If the battery indicator gauge does not light up when you turn on the power:

νCheck the harness connections. Make sure they are tight.

νCheck the circuit breaker. Reset it if necessary.

νCheck the battery connections.

If the above conditions prove normal, you can load test the batteries with a battery load tester. These testers are available at automotive parts stores. Disconnect both batteries before load testing, and follow the directions that came with the load tester. If either one of the batteries fails the load test, replace both of them. If your Jazzy 1113 still does not power up, contact your local Pride provider.

52

www.pridemobility.com

US Jazzy 1113/Rev I/Feb03

Page 52
Image 52
Pride Mobility Jazzy 1113 owner manual Motor Brushes, To inspect or replace the motor brushes, Corrective Maintenance