MAINTENANCE AND TROUBLESHOOTING

Inspect and properly tighten all parts of the exercise cycle regularly. Replace any worn parts immediately.

To clean the exercise cycle, use a damp cloth and a small amount of mild soap. Important: To avoid damage to the console, keep liquids away from the console and keep the console out of direct

sunlight.

PULSE SENSOR TROUBLESHOOTING

Using an adiustable wrench, turn the Left Pedal (20) clockwise and remove it, Next, remove the two M4 x t6mm Screws (40) from the Left Side Shield (15) and the M4 x 12ram Screw (57) from the Side Shield Cover (17). Lift off the Side Shield Cover and the Left Side Shield.

For optimal performance of the pulse sensor, keep the metal contacts clean. The contacts can be cleaned

with a soft cloth--neveruse alcohol, abrasives, or chemicals.

Avoid moving your hands while using the pulse sen- sor. Excessive movement may interfere with heart rate readings. Do not hold the metal contacts too tightly; doing so may interfere with heart rate read- ings. For the most accurate heart rate reading, hold the metal contacts for about 15 seconds.

HOW TO ADJUST THE REED SWITCH

If the console does not display correct feedback, the reed switch should be adjusted. To adjust the reed switch, the left side shield must be removed.

First, remove the five M4 x 25ram Screws (39) from the Right Side Shield (16). Next, remove the M4 x 12mm Screw (57) from the right side of the Side Shield Cover (17).

16

57

39

57

2O

Next, locate the Reed

Switch (46). Turn the left

arm of the

Crank (14) until the

Magnet (58) is aligned with the Reed Switch. Loosen, but do not

remove, the indicated M4 x 16mm Screw (40). Slide the Reed Switch slightly closer to or away from the Magnet. Retighten the Screw. Turn the Crank Arm for a moment. Repeat until the console displays correct feedback. When the Reed Switch is correctly adjust- ed, reattach the left side shield, the side shield cover, and the left pedal.

BATTERY REPLACEMENT

If the console display becomes dim, the batteries should be replaced; most console problems are the result of low batteries. To replace the batteries, refer to assembly step 8 on page 6 and remove the con- sole from the upright. Next, refer to step 7 and insert three batteries into the console. Reattach the console

to the upright. Be careful not to pinch the wires or the cable.

11

Page 11
Image 11
ProForm 831.28317 user manual Maintenance and Troubleshooting, Pulse Sensor Troubleshooting, HOW to Adjust the Reed Switch