Important Information

CAUTION

Do not attempt to modify this product in any way without written authorization from RCA. Unauthorized modification could void the user’s authority to operate this product.

CAUTION

THESE SERVICING INSTRUCTIONS ARE FOR USE BY QUALIFIED SERVICE PERSONNEL ONLY. TO REDUCE THE RISK OF ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING OTHER THAN THAT CONTAINED IN THE OPERATING INSTRUCTIONS UNLESS YOU ARE QUALIFIED TO DO SO.

WARNING

Apparatus shall not be exposed to dripping or splashing and no objects filled with liquids, such as vases, shall not be placed on the apparatus.

CAUTION

When used outside of U.S., other power supply cords may be used if the cord is approved by the local regu- lating agency.

CLEANING AND DISINFECTION

CLEANING : Clean the exterior of this television by removing dust with a lint-free cloth. For further cleaning, use a soft cloth or paper towel dampened with water.

CAUTION : To avoid damage to the surface of the television, do not use abrasive or chemical cleaning agents.

DISINFECTING : Do not immerse this TV, rather clean with a soft damp cloth.

To avoid damage to the surface of the television, test a small portion of the TV’s cabinet with any new disin- fectant to verify that the disinfectant will not discolor or soften the enclosure.

This apparatus with class I construction shall be connected to a main socket outlet with a protective earthing connection.

INSTALLATION PRECAUTIONS

-Any changes or modifications in construction of this television, which are not expressly approved by the party responsible for compliance, could void the user’s authority to operate the equipment.

-Use only a power source from the CPS1000. This TV will operate on DC voltage, 24 volts.

COAX CABLE SPECIFICATION

Coax cable selection is critical when the TV is powered from a central power supply.

Due to long coax cable runs encountered in hospital installations, coax cable employing a solid copper center conductor and copper shield is required. Cable run lengths MUST NOT exceed 150 feet. Required coaxial cable numbers include Alpha 9804C (non-plenum), Belden 9248 (non-Plenum), West Penn 806 (non-Plenum), or West Penn 25806 (Plenum), which have been tested with coax-powered televisions.

WARNING OXYGEN ENVIRONMENT

Do not use in any oxygen tent or oxygen chamber. Such use may cause a fire hazard.

3

Page 3
Image 3
RCA J12H770 manual Cleaning and Disinfection