Problem / possible cause

Action

Component failure (system failure 8, 25).

1.

Power-off the device.

 

2.

Power-on the device again.

LCD: Over Pressure!

The device generates a pressure that is greater than 59 cm H2O. Treatment will be stopped.

1.Power-off the device.

2.Check that the air tubing is connected properly.

3.Power-on the device again.

4.Start Learn Circuit function.

Note: If the alarm activates repeatedly internal components may be defective. Discontinue use and return the device for servicing.

LCD: Blocked Tube!

Air path is blocked.

1.

Check the air path for any blockages.

 

2.

Remove the blockages.

 

3.

If the alarm is not cleared, stop

 

 

treatment.

 

4.

Re-start treatment.

 

 

LCD: High Temperature [10, 11, 12, 23]!

 

 

 

The temperature inside the device is too

Ensure the ambient temperature is within

high. Treatment may lead to stop.

the specified operating range. If the

 

problem persists within the specified

 

operating conditions please return the

 

device for servicing.

 

Contact your clinician.

 

 

 

LCD: High Pressure!

 

 

 

 

 

Therapy pressure exceeds pre-set alarm

1.

Stop treatment.

level.

2.

Re-start treatment.

 

If the problem persists contact your

 

clinician.

 

 

 

LCD: Low Pressure!

 

 

 

 

 

The air tubing is not connected properly.

1.

Check the air circuit integrity and

 

 

reconnect.

 

2.

If the alarm doesn‘t get cleared, stop

 

 

treatment.

 

3.

Re-start treatment.

 

 

 

LCD: Circuit disconnected!

 

 

 

 

 

The air circuit is not connected properly.

1.

Check the air circuit integrity and

 

 

reconnect.

 

2.

If the alarm is not cleared, stop

 

 

treatment.

 

3.

Re-start treatment.

 

 

 

LCD: Low Minute Ventilation!

 

 

 

 

Minute ventilation level has dropped

Contact your clinician.

below the alarm setting level.

 

 

 

 

 