Therapy pressure

IPAP: ± 0.5 cm H2O ± 10% of set pressure (end of

tolerance

inspiration)

 

EPAP/PEEP: ± 0.5 cm H2O ± 4% of set pressure

 

CPAP: ± 0.5 cm H2O ± 10% of set pressure

 

Test condition: T mode, IPAP: 40 cm H2O, EPAP: 2 cm

 

H2O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec,

 

Respiratory Rate: 10 bpm, with ResMed calibration cap.

 

 

Sound pressure level

29 dBA as measured according to ISO 17510 – 1; 2002.

 

32 dBA with uncertainty of 3 dBA as measured according

 

to ISO 17510 – 1:2007.

 

 

Alarm Volume Range

> 45 dBA - <85 dBA at 1 meter (3 steps: low, medium, high)

 

 

Dimensions (L x W x H)

230 mm x 170 mm x 120 mm

 

 

Weight

2.1 kg

 

 

Air outlet

22 mm taper, compatible with ISO 5356-1:2004

 

Anaesthetic & Respiratory Equipment - Conical

 

Connectors

 

 

Pressure measurement

Internally mounted pressure transducer

 

 

Flow measurement

Internally mounted flow transducer

 

 

Power supply

AC 100–240V, 50–60Hz, 2.2 A, max. 65 W

 

 

External DC Power

24 V, 3 A

Supply (isolated)

 

 

 

Internal Battery

Lithium-Ion battery, 14.4 V, 1.6 Ah, 23 Wh

 

Operating hours: 2 h with a new battery under normal

 

conditions (see below).

 

Patient type: home chronic; pressure: IPAP/EPAP 15/5

 

cm H2O; mask type: Ultra Mirage; air tubing: 2 m; leak:

 

0; respiratory rate: 20 bpm; battery capacity: 100%

 

Patient type: hospital acute; pressure: IPAP/EPAP 20/5

 

cm H2O; mask type: Ultra Mirage; air tubing: 2 m; leak:

 

0; respiratory rate: 45 bpm; battery capacity: 100%

 

 

Housing construction

Flame retardant engineering thermoplastic

 

 

Environmental

• Operating temperature: 0°C to 35°C

conditions

• Operating humidity: 10%–95%non-condensing

 

• Storage and transport temperature: -20°C to 60°C

 

(+50°C*)

 

• Storage and transport humidity: 10%–95% non-

 

condensing

 

• Air pressure: 680hPa to 1,100hPa; Altitude: 3,000 m

 

*NONIN XPOD

 

 

Electromagnetic

Product complies with all applicable electromagnetic

compatibility

compatibility requirements (EMC) according to

 

IEC60601-1-2, for residential, commercial, and light

 

industry environments. For further details see “Guidance

 

and manufacturer’s declaration – electromagnetic

 

emissions and immunity” on page 30.

 

 

Air filter

Electro static fibre mesh with TPE frame structure. Bacterial

 

filtration efficiency of 99.540% on area weight 100g/m².

 

 

Air tubing

Flexible plastic, 2 m or 3 m length (22 mm diameter)

 

 

SlimLine air tubing

Flexible plastic, 1.83 m length (15 mm diameter)

 

 

28

Page 33
Image 33
ResMed 100 Therapy pressure, Tolerance, Sound pressure level, Alarm Volume Range, Dimensions L x W x H, Weight Air outlet