Parameter | Description | ||
|
| ||
Learn Circuit | The device is calibrated according to your air tubing system. | ||
| 1. | If in use, turn off the oxygen flow. | |
| 2. | Select the mask type. | |
| 3. | Set up the air circuit including accessories and patient interface. | |
|
| Note: For invasive use, do not connect a catheter mount, | |
|
| tracheostomy tube or HMEF, see “Setting up for invasive use” | |
|
| on page 5). | |
| 4. | Leave the air circuit unobstructed and open to the air. | |
| 5. | Press | to start the Learn Circuit. |
| 6. | Wait for the device to complete its automated tests (<30 sec). | |
|
| The results are displayed when complete. If the circuit | |
|
| configuration has been successfully learnt, displays. If | |
|
| unsuccessful, displays (see “Troubleshooting” on page 22). | |
|
| ||
FiO2 Sensor | The device starts the calibration of the FiO2 monitoring sensor to | ||
Calibration | measure the oxygen concentration of the breathable air. | ||
| 1. | Press | to start the FiO2 sensor calibration. |
| 2. | Wait for the device to complete its calibration. |
The results are displayed when complete.
Note: Turn off the oxygen flow.
Ramp Time | If Max Ramp Time has been set by the clinician, you may select any |
| value up to this time. |
| Options: 0 |
| increments) |
|
|
Setup menu: Configuration Menu
Parameter | Description |
|
|
Language | Sets the display language. |
| Options: Depending on regional configuration |
|
|
Brightness | Sets the LCD backlight brightness. |
| Options: |
|
|
Backlight | Enables the LCD and keypad backlight. |
| If the AUTO setting is selected the backlight turns dark after five |
| minutes without any actions and turns on again if any button is |
| pressed again or an alarm occurs. |
| Options: On, Auto |
|
|
Time format | Sets the time format. |
| Options: 24 hrs,12 hrs |
|
|
Date format | Sets the date format. |
| Options: dd/mm/yyyy, mm/dd/yyyy |
To return to the Options screen press the Return button
English |
Setup menu | 17 |