Problem / possible cause

Solution

The humidifier does not work.

Return the device and the humidifier for

 

servicing.

 

 

The water chamber is empty.

Fill the water chamber of the humidifier.

 

USB stick is not readable or writeable

 

 

The USB stick contains unreadable data,

Consult your clinician.

does not have enough space available, is

 

not compatible with the device.

 

 

 

USB stick is defective.

Replace the USB stick after consulting

 

your clinician.

 

 

FiO2 sensor calibration failure

 

FiO2 sensor is not attached properly.

For the correct attachment of the FiO2

 

sensor, see “Using the FiO2 monitoring

 

sensor” on page 8.

 

 

FiO2 sensor is used or defective.

If the lifetime of the FiO2 sensor has

 

exceeded one year please replace the

FiO2 sensor and start calibration again.

LCD: is displayed in the header.

 

 

Battery is not charging.

• Ensure the ambient temperature

 

 

 

conditions are within the specified

 

 

 

operating range. If the problem persists

 

 

 

within the specified operating conditions

 

 

 

please return the device for servicing.

 

 

 

• Power off the device. Power son the

 

 

 

device again.

 

 

 

Technical specifications

 

 

 

 

 

 

 

Operating pressure

• IPAP: 2 cm H2O to 40 cm H2O (in S, ST, T, PAC mode)

 

 

range

• PS: 0 cm H2O to 38 cm H2O (in S, ST, T, PAC mode)

 

 

 

• EPAP: 2 cm H2O to 25 cm H2O (in S, ST, T, iVAPS, PAC

 

 

 

mode)

 

 

 

• CPAP: 4 cm H2O to 20 cm H2O (in CPAP mode only)

 

 

 

• Min PS: 0 cm H2O to 20 cm H2O (in iVAPS mode)

 

 

 

• Max PS: 0 cm H2O to 30 cm H2O (in iVAPS mode)

 

 

 

Note: iVAPS mode is only available in Stellar 150.

 

 

 

 

 

 

 

Maximum single fault

60 cm H2O (in all modes)

 

 

pressure

 

 

 

 

 

 

 

 

 

Maximum breathing

2 cm H2O at 30 L/min;

 

 

resistance under single

7.2 cm H2O at 60 L/min

 

 

fault

 

 

 

 

 

 

 

 

 

 

 

Maximum flow

> 200 L/min at 20 cm H2O

 

 

Flow accuracy

± 5 L/min or 20% measured value, whichever is greater

 

 

 

 

Test condition: T mode, IPAP: 40 cm H2O, EPAP: 2 cm

 

 

 

H2O, Rise Time: MIN, Fall Time: MIN, Ti: 4.0 sec,

 

 

 

Respiratory Rate: 10 bpm, with ResMed calibration cap.

 

 

 

 

 

English

Technical specifications

27

Page 32
Image 32
ResMed 100 manual Technical specifications