Problem / possible cause

Action

LCD: Keypad Failure!

 

 

 

One of the keys was held down for more

Remove any blockages from the keypad.

than 10 seconds or got stuck.

 

 

LCD: Attention High Temp [42, 43, 44, 45]!

 

 

The temperature inside the device is

Ensure the ambient temperature

high.

conditions are within the specified

 

operating range.

 

 

LCD: Internal Battery Use!

 

 

 

The device is using the internal battery.

Check if the power cord is properly

 

connected to the device if you want to

 

run from mains power.

 

Press the Alarm mute button to cancel

 

the alarm.

 

 

LCD: External DC Power Use!

 

 

 

The device is powered by an external

Check if the AC power cord is properly

battery.

connected to the device if you want to

 

run from mains power.

 

Note: The alarm will be cleared

 

automatically after one minute.

 

 

Other troubleshooting

Problem / possible cause

Solution

No display

 

 

 

Power failure. The device stops delivering

Remove the mask or the catheter mount

air pressure.

from the tracheostomy tube until power

 

is restored.

 

 

Power not connected or device is not

Ensure the power cord is connected and

switched on.

press the switch at the back of the device

 

once.

 

 

Treatment pressure seems low

 

 

 

Ramp time is in use.

Wait for air pressure to build up.

 

 

Air filter is dirty.

Replace air filter.

 

 

Air tubing is kinked or punctured.

Straighten or replace tubing.

 

 

Air tubing is not connected properly.

Connect the air tubing firmly at both

 

ends.

 

 

Mask and headgear not positioned

Adjust position of mask and headgear.

correctly.

 

 

 

Plug(s) missing from access port(s) on

Replace plug(s).

mask.

 

 

 

Pressure required for treatment may

See your clinician to adjust the pressure.

have changed.

 

 

 

There is a large impedance (eg,

Perfom the Learn Circuit function.

antibacterial filter) in the air circuit.

 

 

 

Humidifier control dial set too high,

Turn humidifier control down and empty

resulting in accumulation of water in the

the water from the air tubing.

air tubing.

 

 

 