ResMed 100 manual Vrms

Models: 100

1 42
Download 42 pages 22.66 Kb
Page 36
Image 36

 

 

 

Portable and mobile RF communications

 

 

 

equipment should be used no closer to any part

 

 

 

of the device, including cables, than the

 

 

 

recommended separation distance calculated

 

 

 

from the equation applicable to the frequency of

 

 

 

the transmitter.

 

 

 

Recommended separation distance:

Conducted RF

3 Vrms

3 Vrms

d = 1.17 P

IEC 61000-4-6

150 kHz to 80 MHz

 

 

Radiated RF

3 V/m

3 V/m

d = 1.17 P 80 MHz to 800 MHz

IEC 61000-4-3

80 MHz to 2.5 GHz

 

d = 2.33 P 800 MHz to 2.5 GHz

 

 

 

where P is the maximum output power rating of

 

 

 

the transmitter in watts (W) according to the

 

 

 

transmitter manufacturer and d is the

 

 

 

recommended separation distance in metres (m).

 

 

 

Field strengths from fixed RF transmitters, as

 

 

 

determined by an electromagnetic site survey,a

 

 

 

should be less than the compliance level in each

 

 

 

frequency range.b

 

 

 

Interference may occur in the vicinity of

 

 

 

equipment marked with the following symbol:

 

 

 

 

NOTE 1: Ut is the AC mains voltage prior to application of the test level.

NOTE 2: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 3: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

aField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device.

bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended separation distances between portable and mobile RF communications equipment and the device

The device is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.

 

Separation distance according to frequency of transmitter (m)

Rated maximum output

150 kHz to 80 MHz

 

80 MHz to 800 MHz

 

800 MHz to 2.5 GHz

 

 

power of transmitter (W)

d = 1.17 P

 

d = 1.17 P

 

d = 2.33 P

 

 

 

 

 

 

0.01

0.12

 

0.12

 

0.23

 

 

 

 

 

 

0.1

0.37

 

0.37

 

0.74

 

 

 

 

 

 

1

1.17

 

1.17

 

2.33

 

 

 

 

 

 

10

3.70

 

3.70

 

7.37

 

 

 

 

 

 

100

11.70

 

11.70

 

23.30

 

 

 

 

 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.