ResMed 2011-09, 248551/1 manual Stellar basics, About the control panel

Models: 2011-09 248551/1

1 38
Download 38 pages 47.37 Kb
Page 15
Image 15
Stellar basics

Stellar basics

About the control panel

Mains Power LED

LCD Screen

On when operating on mains power.

External power supply

LED

On when external battery is connected.

Internal battery LED

On when internal battery is in use. Flashes when device is turned off and battery is charging.

Alarm LED About the control panel

Monitoring menu

Setup menu

Info menu

Red or yellow during an alarm or during alarm testing.

Alarm mute button LED

On when Manual backgroundManual background is pressed.

Start/stop button

Key

Start/Stop

Push dial (rotate/

click)

Manual backgroundManual background Therapy LED

On during treatment. Flashes during mask- fitting operation.

Function

Starts or stops treatment.

Extended hold for at least three seconds starts the mask­fitting feature.

Alarm mute

• During therapy: Press once to mute an alarm. Press a

 

 

 

second time to un-mute an alarm. If the problem is still

 

 

 

present, the alarm will sound again after two minutes.

 

 

 

See “Working with alarms” on page 13.

In stand-by mode: Extended hold for at least three seconds starts the alarm LED and alarm buzzer test.

Menu keys

Press the appropriate menu button (Monitoring, Setup, Info)

 

to enter and scroll through the relevant menu.

Push dial

Turning the dial allows you to scroll through the menu and

 

change settings. Pushing the dial allows you to enter into a

 

menu or confirm your choice.