Problem / possible cause

Solution

Pressure required for treatment may

See your clinician to adjust the pressure.

have changed.

 

 

 

There is a large impedance (eg,

Perfom the Learn circuit function.

antibacterial filter) in the air circuit.

 

 

 

Humidifier control dial set too high,

Turn humidifier control down and empty

resulting in accumulation of water in the

the water from the air tubing.

air tubing.

 

 

 

Treatment pressure seems high

 

 

 

Pressure required for treatment may

Consult your clinician.

have changed.

 

 

 

There is a change in the impedance in the

Perform the Learn Circuit function.

circuit configurations.

 

 

The device does not start when you breathe into the mask

 

 

SmartStart/Stop not on.

Consult your clinician.

 

 

Breath is not deep enough to trigger

Take a deep breath in and out through the

SmartStart.

mask.

 

 

There is excessive leak.

Adjust position of mask and headgear.

 

 

Plug(s) missing from port(s) on mask.

Replace plug(s).

 

 

Air tubing is not connected properly.

Connect firmly at both ends.

 

 

Air tubing is kinked or punctured.

Straighten or replace tubing.

 

 

There is a large impedance (eg,

Perform the Learn circuit function.

antibacterial filter) in the air circuit.

 

 

The device does not stop when you remove your mask

 

 

SmartStart/Stop is disabled.

Consult your clinician.

 

 

Use of a full face mask or tracheostomy

SmartStart is disabled if Full face mask or

tube.

Trach is selected as interface.

 

 

Incompatible accessories (eg, humidifier

Use only equipment as recommended

or mask system) with high resistance

and supplied by ResMed.

being used.

 

 

 

High Leak Alarm or Low Min Vent alarms

Consult your clinician.

are set to ON.

 

 

 

Confirm Stopis enabled.

Consult your clinician.

High Leak Alarm is enabled, but alarm does not activate when the mask is removed during treatment

Incompatible air delivery system being

Use only equipment as recommended

used.

and supplied by ResMed.

 

 

Pressure settings are too low for the air

Perform the Learn circuit function to

delivery components being used.

adjust the therapy pressure according to

 

your air tubing system.

 

 

Learn Circuit failed

 

English

The circuit configuration is inappropriate as the impedance detected is too high.

Too many components have been included or the impedance of accessories in use is above ResMed’s recommendation, eg, type of filter, external humidifier, air tubing.

Review the components included in the circuit configuration and adjust as appropriate, then rerun Learn Circuit (see “Setup menu” on page 16).

Troubleshooting 25

Page 30
Image 30
ResMed 248551/1, 2011-09 manual Problem / possible cause Solution, Treatment pressure seems high, Learn Circuit failed